MITS1.5 is designed to simulate high-precision incident fields with user-defined modulation and time sequences as generated by commercial 1.5 Tesla magnetic resonance scanners (also enables worst-case incident field conditions). The system combines cutting-edge technology with accuracy, versatility and user-friendliness. The open user interface allows customized, automated evaluations. MITS1.5 is compatible with the latest draft of the ISO/IEC standard for Active Medical Implant Devices. The system has been rigorously validated. A numerical model of the birdcage is provided. The system is supported by the DASY52NEO scanner, from our sister company SPEAG for rapid and high precision SAR and temperature evaluations. However, MITS systems are standalone products and do not require the DASY52NEO scanner. |
| Application |
Testing compliance of medical implant devices with RF heating guidelines. |
| Operating frequency |
64 MHz |
| Maximum B1 field strength |
> 20 µT |
| Maximum induced E1 |
>> 500 V/m |
| Arbitrary pulse shapes |
10 M samples/sec 14 bit resolution 128 k samples |
| Preprogrammed pulse shapes |
Rectangular Sinc (± 2π) Sinc (± 3π) Gaussian Variable PRR: 1 Hz to 10 kHz Variable duty cycle: 10 to 50 % |
| Length of birdcage |
650 mm |
| Length of bircage with shield | 850 mm |
| Inner diameter of birdcage |
700 mm |
| Weight of birdcage |
39 kg |
| Phantom |
ELIT1.5 for implant lead safety testing compatible with the latest draft of the ISO/IEC standard for Active Implant Medical Devices |
| Robot scanner | Compatible with DASY52NEO from SPEAG |
| Optional | EASY4MRI and SPEAG probes |